Lumex Instruments - Laboratory Analytical Equipment manufacturer Logo
Unit 1-3 7294 Fraserview Place, Mission, B.C., V4S 0A3, Canada V4S 0A3 Unit 1-3 7294 Fraserview Place, Mission, B.C., V4S 0A3 Canada
+1 604-820-8085, +1 604 820-4486, sales@lumexinstruments.com

Lumex Instruments

Insight into mysteries of Nature
  • COVID-19 Detection System
  • Analytical equipment
  • Analytical equipment
  • Analytical equipment
  • Analytical equipment

Products

Lumex Instruments YouTube channel

Determination of arsenic in drinking water by fluorimetric method

Analysis of arsenic in drinking water by fluorimetric method Lumex Instruments 583

INTRODUCTION

Lumex Instruments provides sensitive and selective fluorimetric method for the measurement of mass concentration of arsenic in drinking water samples using the FLUORAT-02 analyzer.

METHOD

The fluorimetric method is based on formation of a complex compound between arsenic (V) and pyrocatechol in weak acid medium, extraction of this compound with chloroform in a form of ion associate with Acridine Yellow G dye followed by measurement of fluorescence intensity of the extract using FLUORAT-02 analyzer displayed as arsenic concentration in mg/l. The result appears on the PC-operated FLUORATE software.

Samples need to be preserved by addition of nitric acid. Carry out the analysis within 24 hours after sampling. Not preserved sample should be analysed within 4 hours after sampling.

File
Determination of arsenic in drinking water by fluorimetric methodDownload application file
EQUIPMENT FOR ANALYSIS
Determination of arsenic in drinking water by fluorimetric method Determination of arsenic in drinking water by fluorimetric method - by
5/ 5stars
Analysis of arsenic in drinking water by fluorimetric method