Portable capillary electrophoresis system Capel-115
Capel-115 is a portable capillary electrophoresis (CE) system intended for determination of a wide range of analytes in various types of samples, including water, soil, food, and fodders using main modes of separations available in capillary electrophoresis.
High performance capillary electrophoresis is based on the differential migration of components of aqueous samples within a narrow fused silica capillary driven by an electric field. Separated solutes are quantitatively detected at the capillary outlet by a high sensitive optical system based on direct or indirect UV absorbance.
- Petrochemical industry
- Environmental analysis
- Food and beverages production
- Animal feed industry
- Portable, light, and field deployable
- Small footprint
- High separation efficiency
- Multiple components determination in one run
- Suitable for various types of analytes from small ions to macromolecules
- Liquid cooling of capillary ensures high reproducibility
- Fast analysis (most analyses take 5–15 minutes)
- Simple sample pre-treatment (just dilute and filter)
- Low analysis cost (nanoliters of reagents, only low-cost consumables are needed)
- Power supply unit equipped with a built-in rechargeable battery and AC power cord
- IQ/OQ protocols
- Capillary electrophoresis system Capel-115
- Spare capillary cassette
- Elforun software package
- Kits & sets for analysis (by request)
- Service agreement (by request)
All Capel-115 CE systems are covered by a 12-month warranty. The information and specifications in this publication are subject to change without notice.
Installation of Capel-115 CE systems can be carried out at a customer's site by our service engineers.
Detection wavelength |
190–400 nm |
Analysis |
Constant voltage, (+/-) 1-25 kV in 1 kV steps |
Injection |
By voltage, (+/-) 1-25 kV in 1 kV steps; |
Rinsing |
By pressure, 500–2000 mbar in 1 mbar steps |
Capillary |
Length 42-100 cm |
Temperature control of capillary |
From 10 °C below ambient temperature (although not lower than +5 °C in absolute value) to +50 °C in absolute value |
Sampler |
Autosampler for 10 vials (standard 1.5 mL) |
Power requirements |
100–240 VAC, 50/60 Hz or built-in battery 15 VDC |
Environmental conditions |
ambient air temperature, 10 – 35 °C |
Time of continuous operation |
5 h |
Dimensions / Weight |
290х385х460 mm, 15 kg |